Programmable Transimpedance Amplifier with Integrated Bandgap Reference for Glucose Concentration Measurement
For glucose electrochemical sensors, a comprehensive electronics interface is designed and constructed in 0.18 um, CMOS process technology, and 1.5 V supply voltage. This interface includes a programmable readout amplifier and bandgap reference voltage potentiostat circuit. The programmable transimpedance amplifier (PTIA), the proposed readout circuit, provides a large dynamic range and low noise. The overall transimpedance increase for the PTIA is 17.3-50.5 kohm. For an input current range of 4.2-180 uA, the PTIA response has a linear output voltage range of 0.55-1.44 V. The output rms noise value is calculated to be 5.101 Vrms, and the overall power consumption of the design is 2.33 mW. The THD percentage spans from 7.6 to 10.2 in the current range mentioned above. All bandgap reference voltage potentiostat measurements are made using the reference potential of 0.6 V. The working electrode was a glassy carbon electrode (GCE) loaded with a CuO/Cu0:76CO2:25O4 (copper cobaltite) coating. An electrochemical glucose sensing setup has been used to measure glucose concentrations between 1 and 10 mM, and an emulated circuit has been used to verify the viability of the proposed glucose sensing design. The suggested glucose sensor architecture has a total size of 0.0684 mm2.
Code (0)
등록된 구현이 없습니다.
Similar Papers 제목 키워드 기반
High-Sensitivity Electric Potential Sensors for Non-Contact Monitoring of Physiological Signals
The paper describes highly-sensitive passive electric potential sensors (EPS) for non-contact detection of multiple biophysical signals, including electrocardiogram (ECG), respiration cycle (RC), and electroencephalogram…
Contact DetectionEEGElectroencephalogram (EEG)SensitivityNoise Analysis of Photonic Modulator Neurons
Neuromorphic photonics relies on efficiently emulating analog neural networks at high speeds. Prior work showed that transducing signals from the optical to the electrical domain and back with transimpedance gain was an …
On-Chip and Off-Chip TIA Amplifiers for Nanopore Signal Readout Design, Performance and Challenges: A Review
Advancements in biomedical research have driven continuous innovations in sensing and diagnostic technologies. Among these, nanopore based single molecule sensing and sequencing is rapidly emerging as a powerful and vers…
BlockingDiagnosticA 125$μ$W 8kS/s Sub-pA Area-Efficient Current Sensing 45nm CMOS ADC for Biosensing
This paper presents a 125$\mu$W, area efficient (0.042mm2) 81dB DR, 8kS/s current sensing ADC in 45nm CMOS capable of sensing sub-pA currents. Our approach combines the transimpedance amplifier (TIA) and ADC into a unifi…
Optimisation of photodetectors design: comparison between Montecarlo and Genetic Algorithms
We present Montecarlo and Genetic Algorithm optimisations applied to the design of photodetectors based on a transimpedance amplifier and a photodiode. The circuit performance is evaluated with a merit function and the s…