Next-generation Surgical Navigation: Marker-less Multi-view 6DoF Pose Estimation of Surgical Instruments
State-of-the-art research of traditional computer vision is increasingly leveraged in the surgical domain. A particular focus in computer-assisted surgery is to replace marker-based tracking systems for instrument localization with pure image-based 6DoF pose estimation using deep-learning methods. However, state-of-the-art single-view pose estimation methods do not yet meet the accuracy required for surgical navigation. In this context, we investigate the benefits of multi-view setups for highly accurate and occlusion-robust 6DoF pose estimation of surgical instruments and derive recommendations for an ideal camera system that addresses the challenges in the operating room. The contributions of this work are threefold. First, we present a multi-camera capture setup consisting of static and head-mounted cameras, which allows us to study the performance of pose estimation methods under various camera configurations. Second, we publish a multi-view RGB-D video dataset of ex-vivo spine surgeries, captured in a surgical wet lab and a real operating theatre and including rich annotations for surgeon, instrument, and patient anatomy. Third, we evaluate three state-of-the-art single-view and multi-view methods for the task of 6DoF pose estimation of surgical instruments and analyze the influence of camera configurations, training data, and occlusions on the pose accuracy and generalization ability. The best method utilizes five cameras in a multi-view pose optimization and achieves an average position and orientation error of 1.01 mm and 0.89\deg for a surgical drill as well as 2.79 mm and 3.33\deg for a screwdriver under optimal conditions. Our results demonstrate that marker-less tracking of surgical instruments is becoming a feasible alternative to existing marker-based systems.
Code (0)
등록된 구현이 없습니다.
Tasks
AnatomyPose EstimationSimilar Papers 제목 키워드 기반
Occlusion-robust Visual Markerless Bone Tracking for Computer-Assisted Orthopaedic Surgery
Conventional computer-assisted orthopaedic navigation systems rely on the tracking of dedicated optical markers for patient poses, which makes the surgical workflow more invasive, tedious, and expensive. Visual tracking …
AnatomyPoint Cloud SegmentationVisual TrackingEnd2Reg: Learning Task-Specific Segmentation for Markerless Registration in Spine Surgery
Intraoperative navigation in spine surgery demands millimeter-level accuracy. Currently, this is achieved through radiation-intensive intraoperative imaging and bone-anchored markers that are invasive and disrupt surgica…
Automatic registration with continuous pose updates for marker-less surgical navigation in spine surgery
Established surgical navigation systems for pedicle screw placement have been proven to be accurate, but still reveal limitations in registration or surgical guidance. Registration of preoperative data to the intraoperat…
AnatomyGPUSurgNavAR: An Augmented Reality Surgical Navigation Framework for Optical See-Through Head Mounted Displays
Augmented reality (AR) devices with head mounted displays (HMDs) facilitate the direct superimposition of 3D preoperative imaging data onto the patient during surgery. To use an HMD-AR device as a stand-alone surgical na…
Effect of Timing Error: A Case Study of Navigation Camera
We focus on the problem of timing errors in navigation camera as a case study in a broader problem of the effect of a timing error in cyber-physical systems. These systems rely on the requirement that certain things happ…
Navigate