Pose Estimation of Periacetabular Osteotomy Fragments with Intraoperative X-Ray Navigation
Objective: State of the art navigation systems for pelvic osteotomies use optical systems with external fiducials. We propose the use of X-Ray navigation for pose estimation of periacetabular fragments without fiducials. Methods: A 2D/3D registration pipeline was developed to recover fragment pose. This pipeline was tested through an extensive simulation study and 6 cadaveric surgeries. Using osteotomy boundaries in the fluoroscopic images, the preoperative plan is refined to more accurately match the intraoperative shape. Results: In simulation, average fragment pose errors were 1.3{\deg}/1.7 mm when the planned fragment matched the intraoperative fragment, 2.2{\deg}/2.1 mm when the plan was not updated to match the true shape, and 1.9{\deg}/2.0 mm when the fragment shape was intraoperatively estimated. In cadaver experiments, the average pose errors were 2.2{\deg}/2.2 mm, 3.8{\deg}/2.5 mm, and 3.5{\deg}/2.2 mm when registering with the actual fragment shape, a preoperative plan, and an intraoperatively refined plan, respectively. Average errors of the lateral center edge angle were less than 2{\deg} for all fragment shapes in simulation and cadaver experiments. Conclusion: The proposed pipeline is capable of accurately reporting femoral head coverage within a range clinically identified for long-term joint survivability. Significance: Human interpretation of fragment pose is challenging and usually restricted to rotation about a single anatomical axis. The proposed pipeline provides an intraoperative estimate of rigid pose with respect to all anatomical axes, is compatible with minimally invasive incisions, and has no dependence on external fiducials.
Code (2)
Tasks
Pose EstimationSimilar Papers 제목 키워드 기반
Fast and Automatic Periacetabular Osteotomy Fragment Pose Estimation Using Intraoperatively Implanted Fiducials and Single-View Fluoroscopy
Accurate and consistent mental interpretation of fluoroscopy to determine the position and orientation of acetabular bone fragments in 3D space is difficult. We propose a computer assisted approach that uses a single flu…
Pose EstimationPatch-Based Image Similarity for Intraoperative 2D/3D Pelvis Registration During Periacetabular Osteotomy
Periacetabular osteotomy is a challenging surgical procedure for treating developmental hip dysplasia, providing greater coverage of the femoral head via relocation of a patient's acetabulum. Since fluoroscopic imaging i…
Electromagnetic Navigation for Femoral Osteotomy Using High-Accuracy X-ray-to-CT Registration
Accurate execution of preoperative plans in corrective femoral osteotomies remains challenging. Current techniques are limited by variable accuracy, invasiveness, and radiation exposure, with free-hand methods and patien…
Real-Time Reconstruction of 3D Bone Models via Very-Low-Dose Protocols
Patient-specific bone models are essential for designing surgical guides and preoperative planning, as they enable the visualization of intricate anatomical structures. However, traditional CT-based approaches for creati…
Knowledge DistillationComparative Analysis of Autonomous Robotic and Manual Techniques for Ultrasonic Sacral Osteotomy: A Preliminary Study
In this paper, we introduce an autonomous Ultrasonic Sacral Osteotomy (USO) robotic system that integrates an ultrasonic osteotome with a seven-degree-of-freedom (DoF) robotic manipulator guided by an optical tracking sy…