Rigid and non-rigid motion compensation in weight-bearing cone-beam CT of the knee using (noisy) inertial measurements

2021-02-24 · Jennifer Maier, Marlies Nitschke, Jang-Hwan Choi, Garry Gold, Rebecca Fahrig, Bjoern M. Eskofier, Andreas Maier

Involuntary subject motion is the main source of artifacts in weight-bearing cone-beam CT of the knee. To achieve image quality for clinical diagnosis, the motion needs to be compensated. We propose to use inertial measurement units (IMUs) attached to the leg for motion estimation. We perform a simulation study using real motion recorded with an optical tracking system. Three IMU-based correction approaches are evaluated, namely rigid motion correction, non-rigid 2D projection deformation and non-rigid 3D dynamic reconstruction. We present an initialization process based on the system geometry. With an IMU noise simulation, we investigate the applicability of the proposed methods in real applications. All proposed IMU-based approaches correct motion at least as good as a state-of-the-art marker-based approach. The structural similarity index and the root mean squared error between motion-free and motion corrected volumes are improved by 24-35% and 78-85%, respectively, compared with the uncorrected case. The noise analysis shows that the noise levels of commercially available IMUs need to be improved by a factor of $10^5$ which is currently only achieved by specialized hardware not robust enough for the application. The presented study confirms the feasibility of this novel approach and defines improvements necessary for a real application.

📄 PDF Abstract BibTeX arXiv:2102.12418

Code (0)

등록된 구현이 없습니다.

Tasks

Dynamic ReconstructionMotion CompensationMotion Estimation

Similar Papers 제목 키워드 기반

Inertial Measurements for Motion Compensation in Weight-bearing Cone-beam CT of the Knee

2020-07-09 · Jennifer Maier, Marlies Nitschke, Jang-Hwan Choi, Garry Gold 외

Involuntary motion during weight-bearing cone-beam computed tomography (CT) scans of the knee causes artifacts in the reconstructed volumes making them unusable for clinical diagnosis. Currently, image-based or marker-ba…

Computed Tomography (CT)Motion Compensation

Multi-Robot Bearing-only Pose Estimation via Angle Rigidity

2026-06-02 · J. Francisco Presenza, Leonardo J. Colombo, Ignacio Mas, Juan I. Giribet arxiv

This letter proposes a novel distributed bearing-based pose estimator for time-varying multi-robot systems. The method uses angles computed from body-frame bearings to estimate the robots' positions in $\mathbb{R}^3$ wit…

Pose Estimation

Retrospective Motion Correction in Gradient Echo MRI by Explicit Motion Estimation Using Deep CNNs

2023-03-30 · Mathias S. Feinler, Bernadette N. Hahn

Magnetic Resonance Imaging allows high resolution data acquisition with the downside of motion sensitivity due to relatively long acquisition times. Even during the acquisition of a single 2D slice, motion can severely c…

compressed sensingMotion CompensationMotion EstimationSensitivity

Angle-based Localization and Rigidity Maintenance Control for Multi-Robot Networks

2026-04-13 · J. Francisco Presenza, Leonardo J. Colombo, Juan I. Giribet, Ignacio Mas arxiv

In this work, we study angle-based localization and rigidity maintenance control for multi-robot networks. First, we establish the relationship between angle rigidity and bearing rigidity considering \textit{directed} se…

Exploring Epipolar Consistency Conditions for Rigid Motion Compensation in In-vivo X-ray Microscopy

2023-03-01 · Mareike Thies, Fabian Wagner, Mingxuan Gu, Siyuan Mei 외

Intravital X-ray microscopy (XRM) in preclinical mouse models is of vital importance for the identification of microscopic structural pathological changes in the bone which are characteristic of osteoporosis. The complex…

Motion Compensation